Single channel infusion
Real-time infusion
Automatic calibration
Automatic power-switch function
Automatic syringe identification
Purge and bolus facility
Dust and water protection exterior
Anti-bolus facility
Memory log function
Qty:
Specifications
| Syringe Size | 10 ml, 20 ml, 30 ml, 50 ml/60 ml |
|---|---|
| Applicable Syringe | Compatible with syringe of any standard |
| VTBI | 1 ml ~ 1000 ml (0.1, 1, 10 ml increments) |
| Flow Rate | 5 ml: 0.1 ml/h ~ 100 ml/h (0.1, 1, 10 ml/h increments) 10 ml: 0.1 ml/h ~ 300 ml/h 20 ml: 0.1 ml/h ~ 600 ml/h 30 ml: 0.1 ml/h ~ 800 ml/h 50/60 ml: 0.1 ml/h ~ 1200 ml/h |
| Bolus Rate | 5 ml: 100 ml/h ( in 0.1, 1, 10 ml/h increments) 10 ml: 300 ml/h 20 ml: 600 ml/h 30 ml: 800 ml/h 50/60 ml: 1200 ml/h |
| Safety Classification | Class II, Type BF |
| Anti-Bolus | Automatic |
| Accuracy | Volume accuracy : within ±2% Mechanical accuracy : within ±1% |
| Infusion Mode | Flow rate: ml/min, ml/h Time-based Body weight: mg/kg/min, mg/kg/h, µg/kg/min, µg/kg/h etc. |
| KVO Rate | 0.1-1 ml/h ( 0.01 ml/h increment ) |
| Alarms | Occlusion, near empty, end program Low battery, end battery, AC power off Motor malfunction, system malfunction Reminder alarm, Pressure sensor error Syringe installation error, Syringe drop off |
| Drug Library | Available |
| Occlusion | HIGH, MED, LOW |
| Memory Log | 50000 events stored |
| Power Supply, AC | 110 / 230 V (optional), 50-60 Hz, 20 VA |
| Battery | 14.8V, Rechargeable |
| Battery Life | 8 hours at 5 ml/h |
| Working Temperature | 5℃ ~ 40℃ |
| Relative Humidity | 20% - 90 % |
| Atmospheric Pressure | 700 hPa ~ 1060 hPa |
| Dimensions | 245 × 120 × 115 mm |
| Weight | 3 kgs |
Features