Hemodialysis Machine HDM-1000E offers a selectable dialysis fluid flow range from 100 mL/min to 800 mL/min. Equipped with a display screen that provides operators with essential treatment data, system status, alerts, and controls. It regulates fluid removal during dialysis to maintain hydration and balance, following preset parameters. Our machine supports ultrafiltration rates from 0 to 4000 mL/h, with an accuracy of ± 1%. It features backup power for the blood pump in case of power failure.

Specifications

Dialysis Fluid Flow Range 100 mL/min to 800 mL/min (Selectable)
Dialysis Fluid Temperature 34.0℃ to 40.0℃ (Selectable)
Dialysis Fluid Conductivity Range 12 to 16 mS/cm at 25℃
Dialysis Fluid Conductivity Accuracy ± 0.1 mS/cm
Ultrafiltration Rate 0 mL/h to 4000 mL/h
Ultrafiltration Accuracy ± 1%
Ultrafiltration Factor unlimited
Ultrafiltration Parameters Displayed UF goal, UF time, UF rate, UF volume
Extracorporeal Arterial Pressure Display Range: -380 mmHg to +400 mmHg
Accuracy: ± 10 mmHg
Resolution: 1 mmHg
Extracorporeal Venous Pressure Display Range: -180mmHg to +600 mmHg
Accuracy: ± 10 mmHg
Resolution: 1 mmHg
Transmembrane Pressure Display Range: -180mmHg to +600 mmHg
Accuracy: ± 20 mmHg
Resolution: 1 mmHg
Blood Flow Range 30 to 600ml/min in 8 mm bloodline system
Blood Flow Accuracy ± 10%
Heparin Pump Syringe Size 10, 20, 30, 50 ml
Heparin Pump Delivery Range 0 ml/h to 10 ml/h
Heparin Pump Bolus Function Maximum 20 ml per bolus
Acid Concentration Range 125 to 150 mmol/L
Acid Concentration Mixing Ratio 1 part acid to 34 parts water 4 (others ratio possible)
Bicarbonate Concentration Range -8 to +8 mmol/L
Bicarbonate Concentration Mixing Ratio 1 part bicarbonate to 27.6 parts water (others ratio possible)
Blood Leak Detector Sensitivity ≤ 0.35 mL blood/min with HCT of 32%
Water Inlet Temperature 5℃ and 32℃
Water Inlet Pressure 1.5 to 6.0 bar
Maximum Drain Height 1m
Supply Pressure 0 to 2 m
Disinfection Program Rinse Temperature: 30℃ to 60℃
Flow rate: 300 mL/min to 800 mL/min
Disinfection Program Hot Rinse Temperature: 30℃ to 85℃
Flow rate: 300 mL/min to 800 mL/min
Disinfection Program with Chemical Temperature: 30℃ to 85℃
Flow rate: 300 mL/min to 800 mL/min
Current 10A
Power Supply AC220V±10%, 47-63HZ
Dimensions (H × W × D) 1380 × 380 × 400 mm
Weight 88 kg

Features

Self-checking function

Double-needle dialysis

Multidetector

Includes heparin, air bubble, and rolling blood pump

Monitors and records various parameters

Automatic disinfection cleaning program

Standard Accessories

Cable

Applications

Hemodialysis Machine perform renal replacement therapy by removing toxins, maintaining fluid balance, and monitoring vital signs of kidney failure patients.

Buy Now :- Hemodialysis Machine HDM-1000E

550    

Dialysis Machine

Medzer Dialysis Machine are engineered for efficient and safe blood purification. They feature an intuitive user interface for easy operation and precise control. Advanced filtration technology maximize toxin removal while maintaining patient comfort. Their compact design save space, while robust construction ensure long-term reliability. They also ensure accurate monitoring of vital parameters throughout the treatment process. Our dialysis machine optimize patient comfort and deliver effective dialysis treatment with consistency.