Medical Equipment Expert
Syringe Size | 10 ml, 20 ml, 30 ml, 50 ml |
VTBI | 0.1 ml ~ 9999 ml |
Flow Rate | 0.1 ~ 400 ml/h (0.1 ml/h step) 0.1 ~ 600 ml/h (0.1 ml/h step) 0.1 ~ 800 ml/h (0.1 ml/h step) 0.1 ~ 800.0 ml/h (0.10 ml/h step) 800 ~ 1200 ml/h (1 ml/h step) |
Purge Rate | 400 ml/h 600 ml/h 800 ml/h 1200 ml/h |
Accuracy | Volume accuracy : within ±2% Mechanical accuracy : within ±1% |
Alarms | Occlusion, near empty, end program Low battery, end battery, AC power off, Power cable disconnected System error, Non-operation timer over |
Occlusion | V: 1200 mmHg ± 200 mmHg H: 800 mmHg ± 200 mmHg M: 500 mmHg±100 mmHg L: 300 mmHg±100 mmHg |
Memory Log | 100 events stored |
Power Supply, AC | AC: 100 V ~ 240 V ± 10%, 50~60Hz±1Hz |
Battery | NIMH battery:DC12V |
Battery Life | 5 hours after 12 hours of charging, at flow rate of 5.0 ml/h |
Working Temperature | 5℃ ~ 40℃ |
Relative Humidity | ≤ 95% |
Dimensions | 295 × 175 × 160 mm |
Weight | 2 kgs |
2024 © All Rights Reserved by Medzer.com