Medical Equipment Expert
Flow Rate | 1ml/h ~ 2000ml/h ; 0.1ml |
VTBI | 0.1ml/h~9999.9ml/h (“0” for turning off the setting function of the volume limit ) Accumulated volume: 0.1ml~9999.9ml |
Bolus | 1000ml/h |
Infusion Mode | Rate mode Time mode Body weight mode Drop counter mode |
Accuracy | +5% (up to +2% after calibration.) |
KVO Rate | 1.0 ml/h ~ 5.0 ml/h |
Alarms | Infusion completion, OCCL. (Upper/Down), low voltage, air bubbles, installation error of infusion pipe, door open, abnormal 1, abnormal 2. |
Rate mode | Rate: 1ml/h ~ 2000.0ml/h |
Time mode | Volume: 0.1ml~999.9 ml Time: 1min ~2000 min |
Body Weight mode | Weight: 0.1 kg ~ 300 kg Drug: 0.1 mg ~ 999.9 mg Volume: 0.1 ml ~ 999.9 ml Dose: 0.1~9999.9 (restricted to unit, drug, solution, weight and infusion pipe) Unit: mg/kg/h, ug/kg/min |
Drop Counter mode | 1 drop/min ~ 666 drops/min |
Occlusion Sensitivity | 400 mmHg ~ 900 mmHg, adjustable, graphical display the pressure. |
Power Supply, AC | AC100V~220V, 50Hz~60Hz |
Battery | Rechargeable Lithium Polymer battery |
Battery Life | 5 hours after being fully recharged (25ml/h) |
Dimensions | 129 × 150 × 172mm |
Weight | <2.2 kgs |