Specifications :

Syringe Size 10 ml, 20 ml, 30 ml, 50 ml, 60 ml
Applicable Syringe Compatible with syringe of any standard
VTBI 0.1 ml ~ 10000 ml  < 100 ml in 0.1 ml increments
Flow Rate Syringe 10 ml: 0.1-420 ml/h
Bolus Rate Syringe 10 ml: 200 ~ 420 ml/h
Syringe 50/60 ml: 800 ~ 1600 ml/h
Anti-Bolus Automatic
Accuracy ±2% (mechanical accuracy ≤ 1%)
Infusion Mode Flow rate: ml/min, ml/h
Body weight: mg/kg/min, mg/kg/h, µg/kg/min, µg/kg/h
KVO Rate 0.1-1 ml/h (in 0.1 ml/h increments)
Alarms Occlusion, near empty, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm, pressure sensor error, syringe installation error, syringe drop off
Drug Library Available
Safety Classification Class II, Type CF
Occlusion Sensitivity LOW, MED or HIGH 
Power Supply, AC AC: 110 V / 230 V (optional), 50 Hz - 60 Hz, 20 VA
Battery 9.6 ± 1.6 V, rechargeable
Battery Life Power saving mode at 5 ml/h
7 hours for double channel
Working Temperature 5℃ ~ 40℃
Relative Humidity 20% ~ 90%
Atmospheric Pressure 860 hpa ~ 1060 hpa
Dimensions 330 × 125 × 225 mm
Weight 5 kgs