| Venous pressure | -50 mmHg ~ +300 mmHg (± 10 mmHg) |
| Arterial pressure | -300 mmHg ~ +300 mmHg (± 10 mmHg) |
| Trans membrane pressure | -100 mmHg ~ +600 mmHg (±20 mmHg) |
| Blood pump | |
| Heparin pump | 0 ml/h ~ 10 ml/h, Accuracy ± 0.1 ml/h |
| Precision | |
| Blood flow range | |
| Dialysate flow range |
300 ml/min ~ 800 ml/min (+10%) 300 ml/min ~ 800 ml/min (-5%) |
| Substitution Fluid Flow | |
| Dialysate temperature | |
| Conductivity | 13 ms/cm ~ 15.5 ms/cm (± 0.1 mS/cm) |
| Air detector | |
| Flow scope | |
| Blood Leak detector | Over 1ml blood per liter dialysate (Flow: 500 ml/min) |
| Hot rinse | ≥ 85℃ |
| Disinfection | Chemical disinfection (citric acid, peracetic acid and oxalic acid) |
| Water supply | |
| Connected load | |
| Power supply | AC 220 V/ 110 V, 50 Hz / 60 Hz, 1500 W |
| Dimension | 370 × 340 × 1570 mm |
| Weight | 100 kgs |
✓ Volume balance control system Enhanced relibility and safety Multi-functional or Individual therapy Air bubble detection system