Medical Equipment Expert
Anesthesia System MAS-1000A supports multiple gas sources including O2, N2O, and AIR, with a pressure range from 280 kPa to 600 kPa. It monitors critical parameters such as frequency, tidal volume, minute volume, and oxygen concentration. It offers various ventilation modes and rapid oxygen supply ranging from 25 to 75 L/min. It features alarms for power failure, low battery, no tidal volume, oxygen concentration, airway pressure, minute volume, continuous pressure, and suffocation. Our system’s built-in backup battery guarantees uninterrupted operation during power outages.
Specifications
Gas Source | O2, N2O, AIR |
---|---|
Pressure | 280 kPa to 600 kPa |
Parameters Monitored | Frequency (Freq): 0 /min to 100 /min Tidal volume (Vt): 0 mL to 2000 mL MV: 0 L/min to 100 L/min Oxygen concentration: 0 % to 100% |
Ventilation Mode | IPPV, V-A/C, P-A/C, V-SIMV, P-SIMV, PCV-PC, PCV-VG, PSV, SIGH, MANUAL |
Flowmeter | O2- 0.1 to 10 L/min N2O- 0.1 to 10 L/min AIR- 0.1 to 10 L/min |
Rapid Oxygen Supply | 25 L/min to 75 L/min |
Tidal Volume (Vt) | 0, 20 mL to 1500 mL |
Frequency (Freq) | 1 /min to 100 /min |
I/E | 4:1 to 1:8 |
PEEP | 0 cmH2O to 30 cmH2O |
Pressure Triggering Sensitivity (PTr) | -20 cmH2O to 0 cmH2O |
Flow Trigger Sensitivity (FTr) | 0.5 L/min to 30 L/min |
Pressure Control (PC) | 5 cmH2O to 70 cmH2O |
SIGH | 0 (off) 1/100 to 5/100 |
Apnea Ventilation | OFF, 5s to 60s |
Pressure Limit | 10 cmH2O to 100 cmH2O |
Waveform Display | P-T (pressure-time) F-T (flow-time) V-T (volume-time) ETCO2-T(ETCO2-time) P-V loop (pressure-volume loop) P-F loop (pressure-flow loop) F-V loop (flow-volume loop) |
AC Power Failure Alarm | Power failure or no connection |
Internal Battery Backup Low Voltage Alarm | < 10.2 ± 0.3 V |
No Tidal Volume | ≤ 5 mL within 6 s |
High Oxygen Concentration Alarm | 19 % to 100 % |
Low Oxygen Concentration Alarm | 18 % to 99 % |
High Airway Pressure Alarm | 20 cmH2O to 100 cmH2O |
Low Airway Pressure Alarm | 0 cmH2O to 20 cmH2O |
High Minute Volume Alarm | Adult- 5 L/min to 20 L/min |
Low Minute Volume Alarm | Paed- 1 L/min to 15 L/min, 0 to 10 L/min |
Continuous Pressure Alarm | PEEP+1.5 kPa over 16 s |
Suffocation Warning | 5 s to 60 s no spontaneous ventilation |
Maximum Limited Pressure | ≤12.5 kPa |
Power Supply | 100 to 240 V 50/60 Hz |
Power Consumption | 80 VA |
Packaging Dimensions (L × W × H) | 920 × 970 × 1380 mm |
Net Weight | 87.5 Kg |
Gross Weight | 155 Kg |
Features
3-level alarm system with audio and visual alerts
Integrated breathing circuit design
Built-in locking mechanisms
Automatic decompression
Multiple waveforms display
10.4-inch TFT LCD screen
Optional Accessories
Applications
2024 © All Rights Reserved by Medzer.com