Catalog for Single Channel ECG Machine SECG-1000A Catalog

Specifications :

Storage Conditions Environment temperature: -10℃ ~ 55℃
Relative humidity: ≤ 95%
Atmospheric pressure: 50 kPa~106 kPa
Operational conditions Environment temperature: +5℃ ~ +35℃
Relative humidity: ≤ 80%
Atmospheric pressure: 86 kPa~106 kPa
Input way Floating and defibrillation protection
Lead Standard 12 leads
Patient leak current < 10 μA
Input impedance ≥5 MΩ
Frequency response 1 Hz ~ 75 Hz
Time constant > 3.2s
CMRR >60 dB
EMG interference filter 25/35 Hz (-3 dB)
Recording way Thermal array printing system
Dimensions of recording paper 50 mm (W) ×20m(L) high-speed thermal paper
Paper speed Auto record: 25 mm/s, 50 mm/s, error: ± 5%
Rhythm record: 25 mm/s, 50 mm/s, error: ±5%
Manual record: 5 mm/s, 6.25 mm/s,10 mm/s,
12.5 mm/s, 25 mm/s, 50 mm/s, error: ±5%.
Sensitivity selections 2.5, 5, 10, 20,40 mm/mV, 5 options, error: ±5%.
Standard sensitivity 10 mm/mV ± 0.2 mm/mV
Measurement parameters 10 mm/mV ± 0.2 mm/mV
Power supply AC:100 ~ 240 V, 50/60 Hz DC:7.4V 3700 mAh rechargeable lithium battery
Product safety type Class I, Type CF defibrillation proof applied part
Enduring polarization voltage ±300 mV
Noise level ≤15 μVp-p
Dimension (L x W x H) 207 mm × 96 mm × 62 mm
Net Weight 0.5 Kg
Features :
  • Single channel
  • Touch screen with Color and large display
  • Light weight, easy to handle
  • Multi-language
  • Built-in rechargeable battery
  • With Auto record set up allows to record format and change leads, measuring and analyzing.
  • High resolution thermal printing array system as high as 75 Hz
  • Auto - Interpretation system and Auto-correct record
  • 50 mm x 20 m high speed thermal paper size.
  • PC software for data transfer and A4 print.
medzer