Venous pressure | 700 mmHg ~ 800 mmHg, Accuracy ± 5 mmHg |
Arterial pressure | 700 mmHg ~+800 mmHg, Accuracy ±5 mmHg |
Trans membrane pressure | 700 mmHg ~ 800 mmHg, Accuracy ±5 mmHg |
Heparin pump | Bolus: 0 ml/time ~ 10 ml/time |
Blood flow range | 0 ml/min ~ 700 ml/min |
Dialysate flow range | 0 ml/min ~ ( 300/ 500 / 1000 ) ml/min (Adjustable) |
Dialysate temperature | 30℃ ~ 41℃, Accuracy ± 0.5℃ |
Conductivity | 0 ~ 25 ms/cm, Accuracy ± 0.1 ms/cm |
Air detector | Detecting ≥ 0.02 mL/single bubble |
Flow scope | UF rate : 0 ~ 6000 ml/h Accuracy: ± 30ml/h or ±1% |
Blood Leak detector | ≥ 0.35 ml/min (HCT 32 %) |
Hot rinse | ≥ 93℃ |
Disinfection | Decalcifiction :Rinse,Citric acid or glacial acetic acid Chemical disinfection:Sodium hypochlorite or peracetic acid Thermal disinfection: Hot water or citric acid |
Water supply | Pressure range: 0 bar ~ 8 bar Temperature range: 0 bar ~ 35 bar Water flow: ≥ 0.3 L/min |
Power supply | AC 220 V / 110 V, 50 Hz / 60 Hz |
Dimension | 1450 × 550 × 460 mm |
Weight | 60 kgs |